Functional Feature : Headset Contact Not Monitored During Active Sessions

Added by Vishal Hawale 20 days ago. Updated about 8 hours ago.

Status:
New
Priority:
Normal
Category:
Functional
Start date:
05/28/2026
Due date:
% Done:

0%

Estimated time:

Description

The headset connection is validated only at the start of a therapy or assessment session. Once the session begins, the headset can be removed from the user's head and placed on a table, yet the session continues and metrics are still displayed as if valid data is being received. This may lead to inaccurate measurements, false-positive results, and reduced confidence in the reported metrics.

Test IDs: 60703, 60705

Suggested Solution:
Implement continuous sensor contact and signal-quality monitoring throughout the session, also an alert must be triggered immediately on disconnection so the user, therapist, or admin can act. A live EEG monitoring feature should be available during active sessions for real-time observation and supervision.