Project

General

Profile

Feature #416

Updated by Mrinmoy Kakoti about 8 hours ago

The headset connection is validated only at the start of a therapy or assessment session. Once the session begins, the headset can be removed from the user's head and placed on a table, yet the session continues and metrics are still displayed as if valid data is being received. This may lead to inaccurate measurements, false-positive results, and reduced confidence in the reported metrics. 

 **Test IDs: 60703, 60705** 

 **Suggested Solution:** 
 Implement continuous sensor contact and signal-quality monitoring throughout the session, also an alert must be triggered immediately on disconnection so the user, therapist, or admin can act. session. A live EEG monitoring feature should be available during active sessions for real-time observation and supervision.

Back