openFunctional Defect: Assessment and Therapy can be conducted without the headset being worn.
Start date:
06/11/2026
Due date:
% Done:
0%
Estimated time:
Description
When only the dongle is powered on and paired with the mobile app (while not connected to a headset or worn by the user), the app incorrectly displays good signal quality (green ticks for both FP1 and FP2).
Under these conditions:
-"Assess Your Wellbeing" can be started and completed.
-Therapy sessions can be conducted and saved.
-No warning is shown regarding sensor contact, signal quality, or headset placement.
Expected Behavior:
The app should verify valid sensor contact and headset placement before allowing assessments or therapies to start. If the headset is not being worn, an appropriate warning should be displayed and the session should be blocked.
Test ID: 60965
Reporting version-1.65